Use the Simple Image Quality Adjustment (SIQA) Tools

Product support for
PrimeLink C9065/C9070 Printer
Article ID
KB0383214
Published
2020-04-10

Simple Image Quality Adjustment (SIQA) Tools Overview

CAUTION:
If using the Optional Xerox Adaptive CMYK Plus Technology:

  • Do not calibrate with Vivid or Fluorescent toner targets or any specialty color toners. Specialty toners do not use a calibration or color profile. Calibration and Profiling should be done on Process color CMYK mode only.

  • If CMYK color accuracy is critical then calibration and profiling should be done after swapping out of a specialty toner set.

Simple Image Quality Adjustment, or SIQA, is a set of maintenance tools that adjust the quality of the printed images generated by the device.
IMPORTANT: SIQA performs three types of adjustments, which must be performed in the following order:

  1. Image Transfer Adjustment: This tool corrects for uneven toner and color shift by creating an Image Transfer Adjustment to the Bias Transfer Roll in the device. The adjustment is created and saved for specific stocks and can be selected for any tray when printing on those stocks. Perform the Image Transfer Adjustment before performing any other image adjustment.

  2. Alignment (Auto Alignment Adjustment): This procedure generates an individual alignment profile for each stock and tray combination you select when performing the procedure. These profiles ensure correct placement of images on the media. You can create a maximum of 50 profiles. Each created profile is then used automatically each time the associated stock is used, ensuring optimal print quality. Perform Alignment Adjustment after you have completed the Image Transfer Adjustment.

  3. Density Uniformity: This procedure adjusts print engine tables, which ensures that ink is deposited uniformly and consistently across the entire surface of a page for each print. Perform the Density Uniformity Adjustment after you have completed the Image Transfer and Alignment Adjustments.

Device calibration with the SIQA tools consists of the following sets of steps:

  1. Print the SIQA targets.

  2. Scan the targets using the Document Feeder or the Platen Glass (document glass).

  3. Save the adjustment data to a file on the device and give the file a unique name.

After the adjustment profile is complete, users can select the profile for print runs, as needed.

Accessing the SIQA Tools

To access the SIQA tools:

  1. At the device control panel, log in as Administrator, press the Machine Status button, then touch the Tools tab. For details, refer to Access the Control Panel or Embedded Web Services (EWS).

  2. In the Features column, scroll down and touch Maintenance.

  3. To access the SIQA tools on the next screen, touch the down arrow.

Image Orientation Definitions

Inboard and outboard are as follows:

  • On the device,inboard andoutboard are defined as follows:

    • Inboard: This term refers to the rear of the device, which is the side of the device that is farthest from you as you are using the device.

    • Outboard: This term refers to the front of the device, which is the side of the device that is nearest to you as you are using the device.

  • On the printed output, as you hold the sheet with the top facing up, inboard refers to the left side of the sheet and outboard refers to the right side of the sheet. The short black line indicates the top of the sheet. The top of the sheet is also referred to as the lead edge.

inboard outboard top-lead edge

Return to Top

Image Transfer Adjustment

The purpose of the Image Transfer Adjustment is to correct for mottle (uneven toner coverage) and color shift (inconsistent color). These image quality issues may occur on heavy-weight stocks. The Image Transfer Adjustment applies only to the media type selected during the procedure. You must perform this adjustment for each media type loaded in the device.

IMPORTANT: It is important to perform Image Transfer Adjustment before performing Auto Alignment Adjustment or Density Uniformity Adjustment.

Adjust Image Transfer

The Image Transfer Adjustment applies only to the paper type selected during the procedure. You must perform this adjustment for each paper type loaded in the device.

To perform a Simple Image Quality Adjustment (SIQA) for image transfer:

  1. At the device control panel, log in as Administrator, then access the SIQA Tools. Refer to Access SIQA.

  2. Touch Adjust Image Transfer.

  3. Touch Paper Type, then select a paper type from the list.

  4. Touch Close.

  5. Load the selected paper type into Tray 5 (bypass).Acceptable paper sizes:

    • 11 x 17 in.

    • 8.5 x 11 in.

    • A3

    • A4

    • SRA3

  6. Touch Print Sample.

  7. Select a paper size.

  8. Select 1 Sided or 2 Sided for the calibration chart.

  9. To print the calibration chart, on the control panel, press the Start button.
    calibration chart sample

  10. Touch Confirm.

  11. Touch Close.

  12. Touch Scan Calibration Chart.

    Note: If you printed a 2-sided sample, scan both sides.

  13. To complete the calibration, follow the instructions on the control panel touch screen.

  14. To exit the scan, touch Confirm.

  15. View the adjustment numbers for Side 1 and Side 2.

  16. To complete and save the calibration, touch Adjust.

  17. To exit the SIQA tool, touch Close, then touch Close again.

  18. To log out of Administrator mode, on the top right corner of the screen, select the Admin icon, then touch Logout.

Adjust Image Transfer Manually

If you are not satisfied with the results achieved after completing the Image Transfer Adjustment procedure, you can run the calibration again using a different row of squares on the calibration chart. You can also change the adjustment values directly.

  1. Select the row of transfer samples that you wish to use for the calibration.

    Note: The key to choosing the best row of samples is to select a row with the fewest white spots in each square.

    pick row from sample

  2. Type the number that represents the desired sample row. The number for the row appears next to the row on the side of the sheet.

    Using the image above as an example, for Side 1 type 1 and for Side 2, type 5

  3. To save the values for image adjustment, touch Adjust.

  4. To exit the SIQA tool, touch Close, then touch Close again.

  5. Touch Close.

  6. To log out of Administrator mode, touch the Admin icon on the top right corner of the screen, then touch Logout.

Return to Top

Auto Alignment Adjustment

The Auto Alignment procedure generates an individual alignment profile, or Type, for each stock/tray combination selected when performing the procedure. These Types ensure that images are placed correctly on the media. You can create up to 50 Types. Each time a stock is used, the Type associated with the stock/tray will be used automatically, ensuring optimal print quality.

IMPORTANT: Before you can perform Auto Alignment Adjustment, you must perform Image Transfer Adjustment. For details, refer to Image Transfer Adjustment.

To perform Auto Alignment Adjustment:

  1. Download the Black Paper PDF file that corresponds to the paper size used in your region.
    The PDF files are located on xerox.com, under the model PrimeLink C9065/C9070 Printer.

    1. Navigate to the Documentation section, then select the associated product (Print Server) from the pull down menu.

    2. Select the required documentation file to download:

      • Xerox SIQA Black Paper for Adjustment A3

      • Xerox SIQA Black Paper for Adjustment Tabloid.

  2. Print a duplex copy of the Black Paper PDF file that corresponds to the paper size used in your region.

    • 11 x 17 in.: SIQA Black Paper for Adjustment-Tabloid.pdf

    • A3: SIQA Black Paper for Adjustment-A3.pdf

  3. At the device control panel, log in as Administrator, then access the SIQA Tools. Refer to Access SIQA.

  4. Touch Alignment Adjustment.

  5. Touch Auto Alignment Adjustment.

  6. Touch Print Calibration Chart.

  7. Select a paper tray with 11 x 17 in. or A3 size paper.

  8. For Coverage 1 and Coverage 2, set the value to 1.

  9. To print the calibration chart, touch Print.
    calibration chart

  10. Touch Confirm.

  11. Touch Document Feeder Scan Precision Adjustment.

  12. To complete the scan using the Document Feeder and Document Glass, follow the instructions on the control panel touch screen.

    1. Position the Black Paper and Calibration Chart as shown on the Black Paper.

    2. Reposition the Calibration Chart as indicated on the control panel touch screen and on the Black Paper, then for each scan, touch Scan.

    3. After you have completed all scans, touch Start.

  13. Touch Confirm.

  14. Touch Print Calibration Chart.

  15. Select a paper tray with the correct paper size.

  16. For Coverage 1 and Coverage 2, select a coverage value. Use the table below as a guide.Table 1. Area Coverage Guide

    Coverage Value

    Area Coverage

    1 - 3

    Low area coverage - text only

    4 - 8

    Medium area coverage - equal mixture of text and graphic images

    9 - 10

    High area coverage - primarily graphic images

  17. Touch Print.

  18. Touch Confirm.

  19. Select a scan method:

    • Touch Scan Chart with Document Feeder: Select this option for a target media size of 11 x 17 in. / A3 or smaller and less than 220 gsm. Accuracy is within 0.2 mm.

    • Touch Scan Chart with Document Glass: Select this option for a target media size of larger than 11 x 17 in. / A3 and more than 220 gsm. Accuracy is within 0.1 mm.

  20. To complete the adjustment, touch Start.

  21. Touch the tray name that displays on the control panel touch screen. This is the paper tray to which the alignment adjustment is applied.

  22. In the Items area, select a type.

  23. Touch Change Settings.

    Note: Selecting Save overwrites the existing settings.

  24. In the Items area, select 1. Name.

  25. To name the Type, touch Change Settings.

    Note: Use a name that identifies the paper type that is loaded in the tray.

  26. To save the settings, touch Save four times.

  27. To log out of administrator mode, select the Admin icon on the top right corner of the screen, then touch Logout.

Setting the Type when Loading Media

Each time you load media, you must associate the paper type with the tray.

To associate the correct Type with the loaded media:

  1. At the device control panel, log in as Administrator, then access the SIQA Tools. Refer to Access SIQA.

  2. Touch Alignment Adjustment.

  3. Touch Manual Alignment Adjustment.

  4. Select the paper tray just loaded.

  5. Touch Change Settings.

  6. Touch the desired Type.

  7. Touch Save.

  8. To save the settings and exit the SIQA tool, touch Close three times.

  9. To log out of Administrator mode, on the top right corner of the screen, select the Admin icon, then touch Logout.

Return to Top

Density Uniformity Adjustment

To perform the Density Uniformity Adjustment:

  1. At the device control panel, log in as Administrator, then access the SIQA Tools. Refer to Access SIQA.

  2. Touch Density Uniformity Adjustment.

  3. Touch Semi Auto Process Using Scanner.

  4. Touch Print Calibration Chart.

  5. Touch Print and follow the instructions on the control panel touch screen to print the two charts.
    print chart for density uniformity adj

  6. Touch Confirm.

  7. Touch Scan Calibration Chart.

  8. Using the instructions on the control panel touch screen, scan the two calibration charts.

  9. To begin the density uniformity adjustment, touch Start.

  10. When the adjustment completes, touch Confirm.

  11. To save the adjustment, touch Save.

  12. To exit the SIQA tool, touch Close three times.

  13. To log out of Administrator mode, on the top right corner of the screen, select the Admin icon, then touch Logout.